Recall

Class II — Beckman Coulter Inc.

Recalling firmBECKMAN COULTER, INC.
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonBeckman Coulter is recalling the Small Form Factor Console PC (SFF PC) (B23083) for use with the DxC 600i analyzer because it may experience a "MFC Exception" error during normal operation of the analyzer system.
ProductUniCel DxC 600i SYNCHRON Access Clinical Analyzer, Catalog No. A25640, Instructions for Use Part No. A93659AB. Product Usage: The DxI 600i systems are in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analytes concentrations found in human body fluids.
Quantity1,519 units total (403 units total)
DistributionWorldwide Distribution - US Nationwide in the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belarus, Bosnia-Herzegovina, Botswana, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, El Salvador, Ethiopia, France, Germany, Gibraltar, Greece, Guam, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhs…
Initiated2014-10-06
Classified2014-11-05
Terminated2015-06-29
Recall numberZ-0176-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Beckman Coulter Inc. (Z-0176-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/69437/

Not open yet

Alerts for BECKMAN COULTER, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.