Class II — Mevion Medical Systems, Inc.
| Recalling firm | Mevion Medical Systems, Inc. |
|---|---|
| Registered facility | MEVION MEDICAL SYSTEMS, INC. (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Software defect that causes an incorrect dose compensation function to be applied to the internal dose ionization chamber when either pressure or temperature sensor malfunction. This could result in an incorrect dose delivery of no more than 5%. |
| Product | MEVION S250, used for proton radiation therapy. |
| Quantity | 1 |
| Distribution | US Distribution in the state of MO |
| Initiated | 2014-11-10 |
| Classified | 2014-12-11 |
| Terminated | 2015-03-18 |
| Recall number | Z-0588-2015 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Mevion Medical Systems, Inc. (Z-0588-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/69778/