RegWardenFDA facility compliance records
Recall

Class II — Carestream Health Inc

Recalling firmCarestream Health, Inc.
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonWhile servicing a unit at a customer site, a field engineer discovered a broken bolt from the carriage assembly. Carestream has identified an issue related to the DRX-Revolution Mobile X-Ray System in which bolts that secure the x-ray tube assembly/boom to the device mounting block may break. It has been determined that this issue is a result of an improperly assembled part which holds the x-ray
ProductDRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243, Catalog Number 1060177 (China) and 1019397 (all other countries) --- Carestream Health, Inc. The DRX-Revolution Mobile X-Ray System is a mobile imaging system that incorporates a self contained x-ray generator, image receptor, imaging display and software for acquiring medical diagnostic images outside of a standard x-ray room. The DRX-Revolution Mobile X-Ray System has a motor for driving the system so that the operator does not have to push the system, but can walk behind the system while it is moving from one location …
Quantity629 units (Domestic: 358 units; Foreign: 271 units)
DistributionWorldwide Distribution: US (nationwide) and countries of: AUSTRALIA, BELGIUM, CANADA, CHILE, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, ITALY, JAPAN, KOREA, KUWAIT, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, and UNITED KINGDOM.
Initiated2014-11-25
Classified2014-12-12
Terminated2016-06-06
Recall numberZ-0800-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Carestream Health Inc (Z-0800-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/69826/

Not open yet

Alerts for Carestream Health, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.