Class II — Horiba Instruments Inc

Recalling firmHORIBA INSTRUMENTS INCORPORATED (DBA) HORIBA Medical
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonHORIBA Medical is informing all HORIBA Medical Lite DM customers using software version 2.0.0.19 and lower of a software bug related to the potential for results received being applied to the wrong order in the Lite DM Worklist. This bug has not been seen in software versions above V2.0.0.19 IMPACT: Potential for results received being applied to the wrong order in the Lite DM Worklist. The is
ProductLite DM (Lite Data Manager) Software Version(s): Versions 2.0.0.19 and lower Product Model #: D00A00224D The Lite DM (Lite Data Manager) is a software application developed by HORIBA Medical intended to receive, store, and manage Startup, Patient, and Quality Control (QC) results from an ABX Micros 60 Hematology Analyzer or ABX Micros CRP 200 Hematology Analyzer. Only one instrument connection is allowed per each Lite DM workstation. The Lite DM application runs on a PC within a Windows Operating System platform. Data transmitted from the instrument to the Lite DM workstation is a Se…
Quantity322 (5 in U.S. Military)
DistributionNationwide within the U.S.
Initiated2014-11-19
Classified2014-12-23
Terminated2015-06-22
Recall numberZ-0856-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Horiba Instruments Inc (Z-0856-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/69847/

Not open yet

Alerts for HORIBA INSTRUMENTS INCORPORATED (DBA) HORIBA Medical

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.