RegWardenFDA facility compliance records
Recall

Class II — Aesculap, Inc.

Recalling firmAESCULAP, INC.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonSome Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in the Instructions for Use (IFU).
ProductMiethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.
Quantity772
DistributionWorldwide Distribution - US Nationwide
Initiated2014-12-17
Classified2015-01-26
Terminated2016-12-13
Recall numberZ-1012-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Aesculap, Inc. (Z-1012-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/70086/

Not open yet

Alerts for AESCULAP, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.