Class II — Zimmer, Inc.
| Recalling firm | ZIMMER, INC. |
|---|---|
| Registered facility | View this company's 10 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Potential for delay in surgery and/or ball bearings to be left in the wound. Repeated cycles of sonic cleaning can break down the ball bearing retention feature (swage) on the shim. To prevent this, TASP shims must be manually cleaned and inspected per product insert, 87-6203-991-22. |
| Product | Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04. Product Usage: The Persona TASPs are sterilizable instruments Intended for multiple uses. The TASP components are used to aid the surgeon in assessing the thickness of the articular surface required. The TASP construct is utilized while conducting a trial range of motion. The TASP consists of 3 parts: a TASP bottom, a TASP shim, and a TASP top. The shims come in varying thicknesses to allow trialing for specific Articular Surface implant thicknesses. |
| Quantity | 67963 |
| Distribution | Worldwide Distribution - USA Nationwide including DC, and PR and the countries of Australia, Austria, Belgium, Canada, Czech Republic, Dominican Republic, France, Germany, India, Iran, Israel, Italy, Japan, Luxembourg, Netherlands, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom. |
| Initiated | 2014-12-11 |
| Classified | 2015-02-04 |
| Terminated | 2015-10-23 |
| Recall number | Z-1052-2015 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Zimmer, Inc. (Z-1052-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/70240/
Alerts for ZIMMER, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.