Class II — Transonic Systems Inc

Recalling firmTRANSONIC SYSTEMS, INC.
Registered facilityTRANSONIC SYSTEMS, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonCustomer unable to obtain measurements due to the device seeing no flow or irrelevant flow condition during dialysis procedures. The user saw a zero or a very low flow value displayed on HD03 and/or no measurements produced in other modes (recirculation; access flow; or cardiac output).
ProductTransonic Hemodialysis Monitor, Model HD03 -- Product Usage: The Hemodialysis Monitor is intended for use by trained medical personnel to measure delivered blood flow, vascular access recirculation, vascular access blood flow and cardiac output on patients receiving hemodialysis treatment.
QuantityDomestic: 21 units, Foreign: 10 units
DistributionWorldwide Distribution - USA Nationwide and the countries of: Australia, Canada, and South Korea
Initiated2015-02-17
Classified2015-03-18
Terminated2016-05-31
Recall numberZ-1295-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Transonic Systems Inc (Z-1295-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/70647/

Not open yet

Alerts for TRANSONIC SYSTEMS, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.