Recall

Class II — CSA Medical

Recalling firmCSA MEDICAL, INC.
Registered facilityCSA Medical (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonTruFreeze Console caused a higher rate of liquid nitrogen (cryogen) to be delivered and may cause: stricture,scarring, bradycardia, or pneumothorax
ProductCSA Medical truFreeze Console- Cryosurgical Unit Cryogenic Surgical Device Model: CC3-01
Quantity82 units
DistributionNationwide
Initiated2015-03-25
Classified2015-04-23
Terminated2015-08-21
Recall numberZ-1513-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)