Class II — Mindray DS USA, Inc. dba Mindray North America
| Recalling firm | Mindray DS USA, Inc. |
|---|---|
| Registered facility | MINDRAY DS USA, Inc. (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the Passport V and Panorama will not be displayed on the Panorama following a restart of either system (due to communication loss, restart, etc.). |
| Product | Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm. Part number 0998-00-0708-01. The Panorama Network includes the Panorama Telemetry System, which acquires and monitors physiological data for ambulating patients within a defined coverage area. |
| Quantity | 4 units |
| Distribution | Worldwide Distribution - US Nationwide in the state of TX. |
| Initiated | 2015-03-12 |
| Classified | 2015-05-26 |
| Terminated | 2017-03-28 |
| Recall number | Z-1675-2015 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Mindray DS USA, Inc. dba Mindray North America (Z-1675-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/70959/
Alerts for Mindray DS USA, Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.