Class II — Philips Medical Systems (Cleveland) Inc

Recalling firmPhilips Medical Systems International BV
Registered facilityView this company's 54 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonIt was noticed that the spectroscopy voxel map did not align with the anatomical images of the phantom. This resulted in incorrect orientation of the Chemical Shift Imaging (CSI) data.
ProductIngenuity TF Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI), 882380, Philips Medical Systems. The device is a diagnostic imaging system for fixed installations that combines Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI).
Quantity10 units
DistributionWorldwide Distribution -- US, including the states of NY, OH, & TX; and, the countries of Finland, Germany, Japan, Netherlands, South Korea, Spain, & Switzerland.
Initiated2015-03-11
Classified2015-05-28
Terminated2018-06-26
Recall numberZ-1682-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Philips Medical Systems (Cleveland) Inc (Z-1682-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/71228/

Not open yet

Alerts for Philips Medical Systems International BV

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.