Class II — Orthofix, Inc
| Recalling firm | ORTHOFIX SRL |
|---|---|
| Registered facility | ORTHOFIX SRL (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | There is a possibility that the thumb button remains in the depressed (unlocked) position allowing the threaded sleeve (drill stop) to unintentionally rotate and change the depth setting. |
| Product | ORTHOFIX Centurion POCT System Drill Guide PN 69-1010 and PN 69-1012. Drill Guides are used to interface with various drill tips during the drilling procedure of bone preparation. |
| Quantity | 36 units |
| Distribution | Nationwide Distribution including CA, VA, FL, ID, CO, OR, PA, OK, SC, and TX. |
| Initiated | 2015-05-11 |
| Classified | 2015-06-22 |
| Terminated | 2016-02-17 |
| Recall number | Z-1819-2015 |
| Initiation | Voluntary: Firm initiated |