Class II — PTW-Freiburg

Recalling firmPTW Freiburg
Registered facilityPTW FREIBURG (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonIn some cases, especially with high dose-per-pulse-rates typical for particle (specifically proton) therapy, measuring errors can result. Individual measurement intervals can be lost that are not visible to the user when there is a high dose-per-pulse rate resulting in a too-low dose being displayed on the monitor. This could result in a radiation over-exposure.
ProductPTW UNIDOS webline: 1) REF T10023, 2) REF T10022 and 3) REF T10023 Firmware version 2.05. Dosimeter for radiation therapy, diagnostic radiology and health physics.
Quantity242 firmwares
DistributionDistributed in CA, MI, MA, NC, NJ, NY, PA, TN, TX, PA & WI.
Initiated2015-05-27
Classified2015-08-21
Terminated2017-03-31
Recall numberZ-2439-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — PTW-Freiburg (Z-2439-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/71395/

Not open yet

Alerts for PTW Freiburg

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.