RegWardenFDA facility compliance records
Recall

Class II — Siemens Medical Solutions USA, Inc.

Recalling firmSIEMENS MEDICAL SOLUTIONS USA, INC.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonPotential measurement error on ACUSON S Family ultrasound system. When repositioning the Virtual Touch IQ region of interest from the original (default) location, the lateral position of the measurement tool may not align with the lateral position of the shear velocity data.
ProductACUSON S2000 and ACUSON S3000 ultrasound systems with the Virtual Touch IQ option; Model numbers: 10041461, 10440017 -S2000 system 10441730 -S3000 system VTIQ Option- 10439521 and 10439522 Radiology: The ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
Quantity629 units
DistributionWorldwide Distribution - US Nationwide and the countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Dominican Rep., France, Germany, Greece, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kenya, Malaysia, Mexico, Nepal, Netherlands, New Zealand, Norway, P.R. China, Philippines, Poland, Portugal, Republic Korea, Romania, Russian F…
Initiated2015-06-25
Classified2015-07-30
Terminated2017-10-19
Recall numberZ-2321-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc. (Z-2321-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/71688/