Class II — Medline Industries Inc
| Recalling firm | Medline Industries, LP |
|---|---|
| Registered facility | View this company's 31 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The print plate artwork for this lot was found to state: Central Line <34 Weeks Kit displaying an incorrect sign in front of the 34. The description for this device should have stated Central Line >34 Weeks Kit. The defect may lead to the product to be used on patient population that is <34 weeks rather than intended population of >34 weeks. |
| Product | Central Line >34 Weeks Kit. Cardiology department. The Central Line >34 Weeks Kit (DYNDC1987A) is intended for use on patients greater than 34 weeks of age. The central venous catheter, also called a central line, is a long, thin, flexible tube used to give medicines, fluids, nutrients, or blood products over a long period of time, usually several weeks or more. A catheter is often inserted in the arm or chest through the skin into a large vein. The catheter is threaded through this vein until it reaches a large vein near the heart. |
| Quantity | 320 units for lot 14DB9277 & 320 units for lot 14PB2878 |
| Distribution | US including PA |
| Initiated | 2015-08-03 |
| Classified | 2015-08-25 |
| Terminated | 2017-01-05 |
| Recall number | Z-2458-2015 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Medline Industries Inc (Z-2458-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/71889/
Alerts for Medline Industries, LP
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.