Class I — SynCardia Systems Inc.

Recalling firmSYNCARDIA SYSTEMS, LLC
Registered facilitySYNCARDIA SYSTEMS, LLC (the firm's only registered facility)
ClassificationClass I
StatusTerminated
ReasonSynCardia Systems, Inc. is recalling Freedom Drivers due to a specific component of the drive mechanism may fail and cause the drive mechanism to stop pumping; resulting in the loss of life-sustaining function.
ProductTemporary Total Artificial Heart (TAH-t) with Freedom Driver System REF 595000-001 Product Usage: Freedom Drivers are components of the SynCardia temporary Total Artificial Heart (TAH-t) System, which is indicated for use in transplant-eligible candidates at risk of imminent death from biventricular failure. Freedom Drivers are intended for use in and out of the hospital.
Quantity56 drivers (21 currently in use on patients)
DistributionWorldwide Distribution -- US Nationwide in the states of AR, NC, CA, OH, PA, TX, MI, and VA; and the countries of Germany, Turkey and France.
Initiated2015-08-06
Classified2015-09-17
Terminated2015-11-04
Recall numberZ-2731-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — SynCardia Systems Inc. (Z-2731-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/71905/

Not open yet

Alerts for SYNCARDIA SYSTEMS, LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.