Class II — Pega Medical Inc.

Recalling firmPega Medical Inc.
Registered facilityORTHOPEDIATRICS CANADA ULC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonSCFE Driver's threaded shaft broke close to the knob when a manual force was applied during the fixation of the screw.
ProductSCFE Driver (SCF-MLD273 & SCF-MLD265) instrument used for the insertion of the Free Gliding SCFE Screw System. This instrument is included in the Free Gliding SCFE Screw System instrument tray. This instrument is marked with the catalogue number and lot number. The Free-Gliding SCFE Screw System is a temporary implant for stabilization of pediatric femoral neck fractures and slipped capital femoral epiphysis (SCFE) in pediatric patients. Orthopedic.
Quantity4 units distributed in the US
DistributionDistributed in the states of FL and TN.
Initiated2015-10-20
Classified2015-11-20
Terminated2016-08-16
Recall numberZ-0324-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Pega Medical Inc. (Z-0324-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/72529/

Not open yet

Alerts for Pega Medical Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.