Class II — DePuy Mitek, Inc., a Johnson & Johnson Co.

Recalling firmDEPUY MITEK, A JOHNSON & JOHNSON COMPANY
Registered facilityDePuy Mitek, Inc. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonMitek identified that on VAPR Tripolar 90 Degree Suction Electrode the ablation and coagulation buttons are colored incorrectly. The wiring for each button function was confirmed as correct and the non-conformance impacted button color only
ProductVAPR® TRIPOLAR 90" Degree Suction Electrodes (Model/Catalog No: 225028)
Quantity127
DistributionWorldwide Distribution - US Distribution to the states of : AL, CA, CT, FL, KY, MN, MS, ND, NJ, NM and TX., and to the countries of: France, Netherlands and Norway.
Initiated2015-11-23
Classified2015-12-23
Terminated2017-04-12
Recall numberZ-0479-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — DePuy Mitek, Inc., a Johnson & Johnson Co. (Z-0479-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/72703/

Not open yet

Alerts for DEPUY MITEK, A JOHNSON & JOHNSON COMPANY

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.