RegWardenFDA facility compliance records
Recall

Class II — Natus Medical Incorporated

Recalling firmNatus Medical Incorporated
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonNatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port detaching from the incubator.
ProductMedix NatalCare ST-LX Intensive Care Incubator. Neonatal incubator.
Quantity75 units (65 units in the US and 10 units in Pakistan) updated in final status report 36 customers with 96 units.
DistributionDistribution US nationwide and Pakistan.
Initiated2016-01-08
Classified2016-04-11
Terminated2017-02-10
Recall numberZ-1363-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Natus Medical Incorporated (Z-1363-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/72894/