Class II — Orthofix, Inc
| Recalling firm | ORTHOFIX SRL |
|---|---|
| Registered facility | ORTHOFIX SRL (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | There is a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In worst case, the unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical intervention. |
| Product | Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512) Intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium). |
| Quantity | 36 units |
| Distribution | Worldwide Distribution -- US, Spain, Italy, and Australia. |
| Initiated | 2016-01-08 |
| Classified | 2016-02-10 |
| Terminated | 2016-11-16 |
| Recall number | Z-0797-2016 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Orthofix, Inc (Z-0797-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/73069/