RegWardenFDA facility compliance records
Recall

Class II — Orthofix, Inc

Recalling firmORTHOFIX SRL
Registered facilityORTHOFIX SRL (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThere is a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In worst case, the unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical intervention.
ProductFirebird Spinal Fixation System Torque Limiting Handle (PN 52-1512) Intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium).
Quantity36 units
DistributionWorldwide Distribution -- US, Spain, Italy, and Australia.
Initiated2016-01-08
Classified2016-02-10
Terminated2016-11-16
Recall numberZ-0797-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Orthofix, Inc (Z-0797-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/73069/