RegWardenFDA facility compliance records
Recall

Class II — Collagen Matrix Inc

Recalling firmCOLLAGEN MATRIX, INC.
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonDuring the 36 month real time shelf life testing of NuOss XC Sinus (BMCU-Umbrella) to verify the 3 year shelf life designated for the device, 3 of 13 products did not pass the recovery test, which requires that the compressed product fully expand within 1 minute of hydration time.
ProductNuOss XC Sinus Reference No 509-3001 (BMCU17, size 17 mm diameter x 10 mm height) Reference No 509-3002 (BMCU22, size 22 mm diameter x 12 m height) Manufacturer:- Collagen Matrix, Distributor:- ACE Surgical Supply Co, Inc, Intended use for dental surgery such as -augmentation or reconstructive treatment of alveoloar ridge -filling of infrabony periodontal defects -filling of defects after root resection, apicoectomy and cystectomy -filling of extraction sockets to enhance preservation of the alveolar ridge -elevation of maxillary sinus floor -filling of periodontal defects in conjun…
Quantity347 units
DistributionUS Distribution to state of: MA.
Initiated2013-05-13
Classified2016-03-11
Terminated2017-06-21
Recall numberZ-1097-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Collagen Matrix Inc (Z-1097-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/73206/

Not open yet