RegWardenFDA facility compliance records
Recall

Class II — Medtronic Inc.

Recalling firmMedtronic, Inc.
Registered facilityView this company's 53 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonMedtronic has identified an issue with the sensitivity of an algorithm used in the Reveal LINQ ICM that may prematurely trigger the Recommended Replacement Time (RRT) alert in some devices. Battery capacity is not affected and the device will continue to support data collection and manual data transmissions. As part of the normal behavior of the device, 30 days after RRT status is reached, Reveal
ProductReveal LINQ insertable Cardiac Monitor, Model LNQ11. Monitors subcutaneous ECG continuously.
Quantity189,009
DistributionWorldwide Distribution-US (nationwide) including DC and the countries of Andorra, Australia, Austria, Belgium, Botswana, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Faroe Islands , Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, Indonesia, Iran, Ireland, Israel, Italy, Kazakhstan, Kenya, Kuwait, Latvia, Lithuania, Luxembourg, Malaysia, …
Initiated2016-02-23
Classified2016-03-25
Terminated2017-09-25
Recall numberZ-1249-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Medtronic Inc. (Z-1249-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/73494/