RegWardenFDA facility compliance records
Recall

Class II — Siemens Medical Solutions USA, Inc

Recalling firmSIEMENS MEDICAL SOLUTIONS USA, INC.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonSiemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system maintenance. The risk of an electrical shock is only present when a service technician ignores several steps listed in the service manual as precautionary actions during device maintenance.
ProductCIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
Quantity4 units
DistributionUS Nationwide Distributed in the states of AR, CA, NE and UT
Initiated2016-02-16
Classified2016-03-31
Terminated2017-11-08
Recall numberZ-1281-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc (Z-1281-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/73546/

Not open yet

Alerts for SIEMENS MEDICAL SOLUTIONS USA, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.