RegWardenFDA facility compliance records
Recall

Class II — Invacare Corporation

Recalling firmInvacare Corporation dba Invacare America
Registered facilityView this company's 7 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe firm recently identified a quality issue with the Junction box used in the Invacare Bariatric Bed. They discovered that the Junction Box may omit sparks which could result in a fire in the presence of flammable materials.
ProductInvacare® Bariatric Bed with mattress and rails , Junction Boxes, Product Usage: The Invacare Bariatric home care bed is intended to provide a rest surface with adjustable height from floor and adjustable surface contours.
Quantity2,702 units
DistributionUS Nationwide Distribution including DC and PR. Internationally Canada.
Initiated2016-03-15
Classified2016-08-11
Terminated2017-09-15
Recall numberZ-2400-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Invacare Corporation (Z-2400-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/73577/

Not open yet

Alerts for Invacare Corporation dba Invacare America

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.