RegWardenFDA facility compliance records
Recall

Class II — CareFusion 303, Inc.

Recalling firmCareFusion 303, INC.
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe Alaris PC units model 8015 may display a system error code 133.6080 due to failure with the super capacitor (C245) at power up on the Alaris PC unit logic boards.
ProductAlaris PC unit, Model 8015 The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System.
Quantity23,397 units
DistributionWorldwide Distribution: US (nationwide) and countries of: Australia, Great Britain and Canada.
Initiated2016-03-24
Classified2016-05-04
Terminated2018-06-29
Recall numberZ-1606-2016
InitiationVoluntary: Firm initiated