Recall

Class II — Stryker Endoscopy

Recalling firmSTRYKER CORP.
Registered facilityView this company's 61 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonWhen operating the device in high flow mode (40L/min, 15 mmHg) and connecting to a house gas inlet source, within seconds of insufflating the flow rate would drop to 0L/min and the red Check Gas Supply symbol would appear on the display, immediately suspending insufflation.
ProductPneumoSure XL High Flow Insufflator and PneumoSure High Flow Insufflator; High pressure units (HPU) (40L and 45L). Model Numbers: 620-040-610, 620-040-611, 620-040-613, 620-040-614, 620-040-600, 620-040-601, 620-040-602, 620-040-604, 105-210-684 The Pneumo Sure XL High Flow Insufflator is a device that serves to create a cavity by insufflating CO2 during diagnostics and/or therapeutic laparoscopy. The insufflator consists of a console, that is intended to be connected to a CO2 gas source, and has a tubeset inserted to transport CO2 gas from the insufflator to the connecting trocar.
Quantity489
DistributionWorldwide distribution. US nationwide including Puerto Rico, China, Hong Kong, India, Japan, Korea, Romania, Vietnam, Singapore, Taiwan, Netherlands, Peru, Australia, Canada, Mexico, Argentina, Chile, Poland, South Africa, and Guatemala.
Initiated2016-03-28
Classified2016-04-25
Terminated2016-12-14
Recall numberZ-1566-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Stryker Endoscopy (Z-1566-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/73674/

Not open yet

Alerts for STRYKER CORP.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.