RegWardenFDA facility compliance records
Recall

Class II — Covidien, LLC

Recalling firmCovidien llc
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonMedtronic is recalling all lots of the Covidien Cytosponge Cell Collection device after two reports of the device detaching from the removal string during withdrawal from the patient's esophagus.
ProductCovidien Cytosponge Cell Collection Kit. Contains Cytosponge Cell Collection Device, Specimen Jar with preservative. Item code CYTO-KIT-R Gastroenterology: Cytosponge Cell Collection Device is indicated for use in the collection and retrieval of surface cells in the esophagus. Currently the device is only available for clinical studies.
Quantity620 total units (260 US; 360 OUS)
DistributionWorldwide Distribution-US Distribution to TN, NC, IL, CA, MN, OR and NY; and countries of: United Kingdom and Iran.
Initiated2016-06-06
Classified2016-07-05
Terminated2017-10-18
Recall numberZ-2122-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Covidien, LLC (Z-2122-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/74300/

Not open yet

Alerts for Covidien llc

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.