RegWardenFDA facility compliance records
Recall

Class II — Sunrise Medical (US) LLC

Recalling firmSunrise Medical International Inc.
Registered facilitySUNRISE MEDICAL INTERNATIONAL INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThere is an observed failure of the backrest bracket which can fatigue over the life time of the product. When this occurs there is the possibility of the bracket fracturing causing the backrest to detach from the chair frame.
ProductQuickie Q7 wheelchair. Model EIR4 Physical Medicine. The Quickie Q7 series of wheelchair's intended use is to provide mobility to persons limited to a sitting position.
Quantity12,142 devices in US, 1,710 devices internationally.
DistributionUS and Mexico, Japan, Chile, Kuwait, Columbia, Hong Kong, China, Uruguay, South Africa, Costa Rica, South Korea, Singapore, New Zealand, El Salvador, Bahrain, Uruguay, Lebanon, United Arab Emirates, Brazil, Argentina, Netherlands Antilles, British Virgin Islands,
Initiated2016-07-01
Classified2016-07-25
Terminated2017-09-11
Recall numberZ-2290-2016
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Sunrise Medical (US) LLC (Z-2290-2016).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/74596/