RegWardenFDA facility compliance records
Recall

Class II — ConMed Corporation

Recalling firmCONMED CORPORATION
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe packaging seal may contain a crease. The crease may result in an open channel. These channels may compromise sterility of the product.
ProductCORE® Suction Irrigation Handpiece without Probe for Single Solution or Dual Bags, Catalog Number CD8302 The CORE Trumpet (CD81XX, CD83XX, and CD84XX) has applications in gynecological laparoscopy, laparoscopic cholecystectomy and other laparoscopic procedures.
QuantityN/A
DistributionWorldwide Distribution - US (Nationwide) Internationally: to Australia, Austria, Uruguay, Denmark, Germany, Ireland, Italy, United Kingdom, Guatemala, Czech Republic & Canada.
Initiated2016-06-27
Classified2016-10-18
Terminated2018-06-07
Recall numberZ-0133-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — ConMed Corporation (Z-0133-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/74774/