Class II — Handicare USA, Inc.
| Recalling firm | Handicare USA, Inc. |
|---|---|
| Registered facility | Handicare USA, Inc. (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | During the period of April 2009 May 2011, Handicare AB used a wing handle in which the screw and handle, after repeated tightening, may slide apart. This causes a small amount of play in the attachment of the ladder, which in turn results in high load at the attachment holes of the ladder. Eventually, this may cause mechanical fatigue in the material of the ladder. |
| Product | ReTurn 7500/7500 Aid, transfer |
| Quantity | 29 units , plus 47 wing handles were shipped as spare parts |
| Distribution | US Distribution to the states of : NY, NJ, NV, NE, IL, CT, PA and ME |
| Initiated | 2014-04-15 |
| Classified | 2016-10-24 |
| Terminated | 2016-12-13 |
| Recall number | Z-0303-2017 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Handicare USA, Inc. (Z-0303-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/75404/
Alerts for Handicare USA, Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.