Class III — Angiodynamics, Inc.
| Recalling firm | ANGIODYNAMICS, INC |
|---|---|
| Registered facility | View this company's 4 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class III |
| Status | Terminated |
| Reason | Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits. |
| Product | ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T ECHO B PG, Item Number H787065970465, Catalog No.REF 06597046, 2) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T ECHO S PG, Item Number H787065970485, Catalog No. REF 06597048 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick |
| Quantity | 80 boxes |
| Distribution | Nationwide Distribution to AL, AR, CA, FL, GA, IL, KY, LA, MI, MN, MO, NC, OH, OK, SC, NY, ME, PA, MD, VA, WA, WI, WV, TN, and TX |
| Initiated | 2016-09-28 |
| Classified | 2016-12-16 |
| Terminated | 2018-06-18 |
| Recall number | Z-0773-2017 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class III recall — Angiodynamics, Inc. (Z-0773-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/75420/