Class II — Siemens Healthcare Diagnostics, Inc.

Recalling firmSIEMENS HEALTHCARE DIAGNOSTICS INC.
Registered facilityView this company's 12 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonSiemens has confirmed that the DBIL_2 Assigned Calibrator Values (ACV) currently assigned to the affected lots may cause, on average, a positive proportional bias of approximately 41% on patient samples compared to the reference method for Direct Bilirubin measurement.
ProductAdvia Chemistry Calibrator For in vitro diagnostic use in the calibration of ADVIA® Chemistry systems for the following methods: Albumin (ALB, ALB_c), Direct Bilirubin (DBIL_2), Total Bilirubin (TBIL_2), Calcium (CA, CA_c, CA_2, CA_2c), Cholesterol (CHOL), Creatinine (CREA, CREA_2, CRE_2c, ECRE_2), Glucose Hexakinase (GLUH), Glucose Oxidase (GLUO), Iron (IRON_2), Lithium (LITH), Magnesium (MG, MG_c), Phosphorus - inorganic (IP, IP_c), Total Protein (TP, TP_c), Triglycerides (TRIG), Urea Nitrogen (UN, UN_c), Uric Acid (UA, UA_c)
Quantity10,614 units
DistributionWorldwide Distribution - US (Nationwide) Internationally to Canada, Mexico, Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, Chile, Cyprus, Denmark, Egypt, Ecuador, Finland, France, Germany, Great Britain, Hong Kong, Hungary, Italy, Japan, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Poland, Portugal, Russian Fed. Slovakia, Slovenia, Spain, Swed…
Initiated2016-10-07
Classified2016-11-16
Terminated2018-05-10
Recall numberZ-0589-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Healthcare Diagnostics, Inc. (Z-0589-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/75487/

Not open yet

Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.