Recall

Class I — SynCardia Systems Inc.

Recalling firmSYNCARDIA SYSTEMS, LLC
Registered facilitySYNCARDIA SYSTEMS, LLC (the firm's only registered facility)
ClassificationClass I
StatusTerminated
ReasonThe Main Printed Circuit Board Assembly (PCBA) of the affected Freedom Drivers may fail and cause the Freedom Driver to stop functioning without visual or audible alarms, resulting in the loss of life-sustaining function.
Product5000 Series Freedom Drivers. Freedom Driver System for Temporary Total Artificial Heart (TAH-t). Part number 595000-001
Quantity5
DistributionUS distribution to Virginia and Arizona.
Initiated2016-10-21
Classified2016-12-08
Terminated2017-02-21
Recall numberZ-0659-2017
InitiationVoluntary: Firm initiated