Class II — Siemens Medical Solutions USA, Inc
| Recalling firm | SIEMENS MEDICAL SOLUTIONS USA, INC. |
|---|---|
| Registered facility | View this company's 9 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems. |
| Product | SOMATOM Perspective, Computed tomography x-ray system |
| Quantity | 13 systems |
| Distribution | Distributed to: IL IL GA SC SD OK NY MI IN MD AR |
| Initiated | 2016-11-22 |
| Classified | 2016-12-23 |
| Terminated | 2017-05-30 |
| Recall number | Z-0921-2017 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)