Class II — The Binding Site Group, Ltd.
| Recalling firm | THE BINDING SITE, INC. |
|---|---|
| Registered facility | THE BINDING SITE, INC. (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | A change in the calibration curve causing an increase in false prozone flags. |
| Product | Freelite Human Lambda Free Kit for use on SPAPLUS Product Code: LK018.S, LK018.10S Freelite aids the detection and monitoring of Multiple Myeloma and related diseases |
| Quantity | 3,088 |
| Distribution | Worldwide Distribution - U.S. (nationwide) and Internationally to the following; Canada, Australia. |
| Initiated | 2016-10-28 |
| Classified | 2017-01-19 |
| Terminated | 2017-07-27 |
| Recall number | Z-1053-2017 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — The Binding Site Group, Ltd. (Z-1053-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/76050/