RegWardenFDA facility compliance records
Recall

Class II — Becton Dickinson & Company

Recalling firmBecton Dickinson and Company
Registered facilityView this company's 36 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonA portion of this lot of blood collection tubes was manufactured with approximately twice the expected amount of buffered sodium citrate additive, an anticoagulant.
ProductBD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (4.5 ml x 13 x 75 mm) Catalog Number: 369714 Product Usage: The BD Vacutainer Safety Coagulation tube is an evacuated blood collection tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations
Quantity1,488,800 tubes
DistributionWorldwide Distribution - US Nationwide and the countries of Canada, Mexico, Indonesia
Initiated2017-01-16
Classified2017-02-10
Terminated2018-02-07
Recall numberZ-1189-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Becton Dickinson & Company (Z-1189-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/76295/