RegWardenFDA facility compliance records
Recall

Class II — Smith & Nephew, Inc.

Recalling firmSmith & Nephew, Inc.
Registered facilityView this company's 15 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe single use devices are provided sterile and do not have an expiration date on the label.
ProductSmith & Nephew Dyonics Saw Blade, Short, Product Number 3703
Quantity4500
DistributionNationwide Foreign: Brazil, Canada, Chile , China, Denmark, Hong Kong, India, Mexico, Peru, Singapore, Venezuela
Initiated2017-02-02
Classified2017-02-22
Terminated2019-09-20
Recall numberZ-1241-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Smith & Nephew, Inc. (Z-1241-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/76414/

Not open yet

Alerts for Smith & Nephew, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.