Recall

Class II — Medtronic Inc.

Recalling firmMedtronic
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonMedtronic has identified a software anomaly that can prevent the internal battery of the pump from charging. When the software anomaly occurs it leads to battery depletion and pump shutdown.
ProductMiniMed 620G and MiniMed 640G Insulin Pumps with Software Version 2.6 and below are affected. Catalog Numbers for these insulin pumps are MMT-1510, MMT-1710, MMT-1711, MMT-1712.
Quantity42,215 units
DistributionEurope, Australia, Japan, Singapore
Initiated2017-01-31
Classified2017-04-28
Terminated2017-05-04
Recall numberZ-1890-2017
InitiationVoluntary: Firm initiated