RegWardenFDA facility compliance records
Recall

Class II — Encore Medical, Lp

Recalling firmENCORE MEDICAL, L.P.
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonDuring the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.
ProductTuron Impaction Fixture
Quantity297 units
DistributionUS, South Korea, Australia, Canada, United Kingdom/Ireland, Germany
Initiated2017-02-09
Classified2017-03-08
Terminated2017-10-20
Recall numberZ-1412-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Encore Medical, Lp (Z-1412-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/76456/