RegWardenFDA facility compliance records
Recall

Class II — United Orthopedic Corporation

Recalling firmUNITED ORTHOPEDIC CORPORATION
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe locking lever pin component of the femoral driver may become deformed and break, resulting in malfunction during surgery and causing the device to become inoperable.
ProductUOC Femoral Driver, Product number PE 93045101, orthopedic manual surgical instrument/accessory for use during surgery to implant femoral components and femoral trial, and to remove femoral trial.
Quantity33 units
DistributionWorldwide Distribution - US including State of CA and Internationally to Pakistan, Spain, and Taiwan.
Initiated2017-02-10
Classified2017-03-29
Terminated2017-06-28
Recall numberZ-1692-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — United Orthopedic Corporation (Z-1692-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/76540/

Not open yet

Alerts for UNITED ORTHOPEDIC CORPORATION

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.