RegWardenFDA facility compliance records
Recall

Class II — Angiodynamics Inc. (Navilyst Medical Inc.)

Recalling firmANGIODYNAMICS, INC
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonNavilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFlo PICC, conducted this recall to the end user level based on information received from Greatbatch Medical, the manufacturer of the ViaPeel PTFE Peelable Introducer. Greatbatch Medical determined that the products listed in their 11/11/2016 Recall Notification have the potential for the handles to detach from the sheath during use. NMI has confirmed that the affected sheaths, Greatbatch Model Number 10890-006, had been included in packaged Xcela and BioFlo PICC Kits.
ProductXCELA PASV 5F DL 55CM IR-145 KIT PG, UPN H965457330, Catalog No. 45-733 The Xcela with PASV, BioFlo PICC with ENDEXO Technology, and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.
Quantity4591 units in total
DistributionWorldwide Distribution: US (Nationwide) and countries of:; Canada, Great Britain, Netherlands, South Australia and Spain.
Initiated2016-11-28
Classified2017-12-13
Terminated2019-06-12
Recall numberZ-0243-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Angiodynamics Inc. (Navilyst Medical Inc.) (Z-0243-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/76570/

Not open yet

Alerts for ANGIODYNAMICS, INC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.