Class II — Strata Skin Sciences, Inc.

Recalling firmSTRATA Skin Sciences, INC.
Registered facilityStrata Skin Sciences, inc. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonWhen the Liquid Guide ( LLG) is not used according to the operator manual, the switch control wires may become loose and shorted. Shorted wires may lead to unintentional laser firing.
ProductXTRAC Excimer Laser, Model No. AL8000, AL10000 The intended use is targeted UVB phototherapy for treatment of the skin conditions including psoriasis, vitiligo, atopic dermatitis, and leukoderma.
QuantityUS - 1000 , Foreign - 612
DistributionUSA (nationwide) Distribution
Initiated2017-01-23
Classified2017-03-27
Terminated2017-06-05
Recall numberZ-1455-2017
InitiationFDA Mandated