RegWardenFDA facility compliance records
Recall

Class II — Intuitive Surgical, Inc.

Recalling firmIntuitive Surgical, Inc.
Registered facilityView this company's 41 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonIntuitive Surgical has become aware that in specific scenarios with system software P6 and the da Vinci Xi Suction Irrigator (PN 480299-03), users can experience unexpected motion of a system arm.
Productda Vinci Xi EndoWrist Suction Irrigator, 8 mm instrument; PN 480299-03; and SOFTWARE, EMBEDDED RLS, IS4000, A70_P6_B440; PN 610092-440. General and Plastic Surgery: The EndoWrist® Suction Irrigator is designed to be used in conjunction with an Intuitive Surgical da Vinci Surgical System and compatible suction and irrigation sources and tubing sets for delivering fluid to the surgical site and for evacuation and aspiration of fluids. The instrument may also be used for retraction and blunt dissection of tissue. The instrument tip is blunt and intended to contact tissue.
Quantity15 x 6 packs (90 units)
DistributionUS Only - one location each in AL, CO, KS, NV, NY, and 2 in TX
Initiated2017-03-31
Classified2017-04-13
Terminated2017-07-28
Recall numberZ-1819-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Intuitive Surgical, Inc. (Z-1819-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/76911/

Not open yet

Alerts for Intuitive Surgical, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.