Class II — Luminex Corporation
| Recalling firm | Luminex Corporation |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The hand-held barcode scanner that is used as an accessory to enter ARIES cassette and specimen identifications on the ARIES Systems populates patient sample IDs with a previously scanned sample ID when entering orders into an ARIES instrument. |
| Product | ARIES System and ARIES M1 System |
| Quantity | 15 units |
| Distribution | Worldwide Distribution - US Distribution and to the countries of Finland and China. |
| Initiated | 2017-04-19 |
| Classified | 2017-05-16 |
| Terminated | 2017-10-20 |
| Recall number | Z-2094-2017 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Luminex Corporation (Z-2094-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/77052/