Class II — Novo Nordisk Inc
| Recalling firm | NOVO NORDISK INC. |
|---|---|
| Registered facility | View this company's 5 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | A small number of batches may crack or break if exposed to certain chemicals, like certain cleaning agents. |
| Product | Dial-A-Dose Insulin Delivery Device (Pen-Injector) cartridge holders Product Usage: The NovoPen Echo® is a re-usable pen injector for single patient use by diabetics for the self injection of a desired dose of insulin. The pen injector uses Penfill 3 mL cartridge of NovoLog®, 100 units/mL (Ux100) [insulin aspart (rDNA origin) injection] and a single use detachable and disposable pen needle (supplied separately). The pen injector allows the user to dial the desired dose from 0.5 to 30 units in 0.5 increments. |
| Quantity | 16,557 units |
| Distribution | Nationwide distribution |
| Initiated | 2017-07-05 |
| Classified | 2017-12-07 |
| Terminated | 2019-04-09 |
| Recall number | Z-0206-2018 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Novo Nordisk Inc (Z-0206-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/77704/
Alerts for NOVO NORDISK INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.