| Recalling firm | FIREFLY MEDICAL, INC. |
|---|---|
| Registered facility | FIREFLY MEDICAL, INC. (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The dowel pins (P/N 16082) that fit in the left and right leg pivot castings (P/Ns 16049/16056) can rotate and gradually maneuver out of the leg casting after several hundred deployment and folding cycles (designed for easy storage) of the IVEA unit. |
| Product | IVEA Product Part Number 16100 Model 500A |
| Quantity | 570 units |
| Distribution | United States, Australia, Belgium, Canada, China, Germany, Hong Kong, Indonesia, Israel, Netherlands, Singapore, UAE |
| Initiated | 2017-08-24 |
| Classified | 2017-09-29 |
| Terminated | 2019-11-01 |
| Recall number | Z-3239-2017 |
| Initiation | Voluntary: Firm initiated |