Class II — 3M Company - Health Care Business
| Recalling firm | 3M COMPANY |
|---|---|
| Registered facility | View this company's 13 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower than the patients actual temperature. |
| Product | 3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000) |
| Quantity | 153 cases (3825 sensors) |
| Distribution | CA, MI, IL, MO, MN, GA, and Canada, Switzerland |
| Initiated | 2017-10-17 |
| Classified | 2017-11-18 |
| Terminated | 2020-11-03 |
| Recall number | Z-0114-2018 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — 3M Company - Health Care Business (Z-0114-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/78327/