RegWardenFDA facility compliance records
Recall

Class I — Spacelabs Healthcare, Ltd.

Recalling firmSPACELABS HEALTHCARE, INC.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusTerminated
ReasonArkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display unit that resulted in a blank screen without audible or visible alarms.
ProductArkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician
Quantity110 units
DistributionNationwide Distribution to the states of NM, WY, FL, NC, MS, CO, CT, ME, and AL.
Initiated2017-10-11
Classified2017-11-21
Terminated2018-03-12
Recall numberZ-0072-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Spacelabs Healthcare, Ltd. (Z-0072-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/78330/