Class II — Horiba Instruments, Inc dba Horiba Medical

Recalling firmHORIBA INSTRUMENTS INCORPORATED (DBA) HORIBA Medical
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.
Product15 ML REAGENT CUP, B1037307, SAP 1221037307 for use with ABX PENTRA 400 / PENTRA C400
Quantity10 bags of 100 units
DistributionNationwide
Initiated2017-11-21
Classified2018-01-31
Terminated2019-11-04
Recall numberZ-0494-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Horiba Instruments, Inc dba Horiba Medical (Z-0494-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/78692/

Not open yet

Alerts for HORIBA INSTRUMENTS INCORPORATED (DBA) HORIBA Medical

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.