Class II — Beckman Coulter Inc.
| Recalling firm | BECKMAN COULTER, INC. |
|---|---|
| Registered facility | View this company's 22 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Unpaired lots of AU lyophilized chemistry calibrator level 1 and 2 have been distributed. Calibrator level 1 and 2 must be paired when used for calibration and cannot be used in any other combination. If lot numbers are not correctly paired, calibration can be run using incorrect set points which may generate erroneous results. QC failures could occur if incorrect pairs of calibrators are used. |
| Product | Lyophilized Chemistry Calibrator Level 1 Product Usage: Beckman Coulter Chemistry Calibrators are intended for use when calibrating methods run on the Beckman Coulter AU series of chemistry analyzers. Beckman Coulter Chemistry Calibrators are lyophilized, human serum based products formulated for use as a reference material when calibrating Beckman Coulter AU clinical chemistry system assays. The calibrators are used to calibrate the following assays: Albumin (ALB), Bicarbonate (CO2), Direct Bilirubin (DBIL), Total Bilirubin (TBIL), Calcium Arsenazo (CALA), Calcium (CAL oCPC), Cholesterol… |
| Quantity | unknown |
| Distribution | Worldwide Distribution in the states of AL, AK, AZ, AK, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IO, KS, KY, LA, MA, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NY, NM, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VI, WA, WV, WI, and WY and the countries of Brazil, Canada, Costa Rica, El Salvador, Guam, Paraguay, and Peru |
| Initiated | 2017-08-24 |
| Classified | 2018-02-13 |
| Terminated | 2020-08-28 |
| Recall number | Z-0605-2018 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Beckman Coulter Inc. (Z-0605-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/78699/