Class II — Medistim Asa

Recalling firmMedistim ASA
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonA system component overheated causing melting of the connector and emission of smoke through the systems loudspeaker opening.
ProductMEDISTIM VERIQ SYSTEM, MODEL VQ4122 An intraoperative diagnostic system that utilizes ultrasonography to guide surgeons to successfully plan and accomplish surgical interventions.
Quantity2
DistributionDistributed to New York.
Initiated2017-04-25
Classified2018-02-26
Terminated2024-05-03
Recall numberZ-0719-2018
InitiationVoluntary: Firm initiated